DVM851
01.08.2012
6
condensation is to be expected.(home and office environments fall under this
category)
tive pollution occurs, or dry nonconductive pollution occurs that becomes
conductive due to condensation that is to be expected.
(industrial environments and environments exposed to outside air
contact with precipitation)
he pollution generates persistent conductivity caused by conductive dust or by rain
or snow. (exposed outdoor environments and environments where high humidity
levels or high concentrations of fine particles occur)
Warning:
This device was designed in accordance with EN 61010-1 pollution degree 2
. This implies that certain
restrictions in use apply that are related to pollution which can occur within the environment of use.
Refer to the table above.
This device is only suitable for measurements in Pol
lution degree class 2 environments.
9. Specifications
This device is not calibrated when purchased!
Regulations concerning environment of use:
•
Use this meter only for measurements in CAT I and CAT II environments (see §7)
• Use this meter only in a pollution degree 2 environment (see §8)
Ideal working conditions include:
• Temperature: 18 to 28°C (64°F to 82°F)
• relative humidity: max. 80%
• altitude: max. 2000m (6560ft)
slope integration A/D converter
Only figure "1" on the display
" displayed for negative polarity
9.1 DC VOLTAGE
Do not measure circuits that may contain voltages > 600V
±
±
±
9.2 DC CURRENT
Do not measure current in circuits with voltages > 250V
±
±
±
Overload protection by fuse: 200mA range F 500mA/250V, 10A range F10A/
©Velleman nv
condensation is to be expected.(home and office environments fall under this
tive pollution occurs, or dry nonconductive pollution occurs that becomes
(industrial environments and environments exposed to outside air
- but not in
he pollution generates persistent conductivity caused by conductive dust or by rain
or snow. (exposed outdoor environments and environments where high humidity
. This implies that certain
restrictions in use apply that are related to pollution which can occur within the environment of use.
lution degree class 2 environments.
Use this meter only for measurements in CAT I and CAT II environments (see §7)
slope integration A/D converter
Only figure "1" on the display
" displayed for negative polarity
±
± 2 digits
±
± 2 digits
±
±